Open procedure |
 |
Restricted procedure |
b) The products involved are manufactured purely for the purpose of research, experiment, study or
development under the conditions stated in the Directive (for supplies only).
c) The works/goods/services can be provided only by a particular tenderer for reasons that are:
 |
Technical |
 |
Artistic |
 |
Connected with the protection of exclusive rights |
d) Extreme urgency brought about by events unforeseeable by the contracting authority and in
accordance with the strict conditions stated in the Directive.
e) Additional works/deliveries/services are ordered under the strict conditions stated in the Directive.
f) New works/services, constituting a repetition of existing works/services and ordered in accordance
with the strict conditions stated in the Directive.
g)Service contract awarded to the successful candidate or one of them after a design contest.
h) For supplies quoted and purchased on a commodity market.
i) For the purchase of supplies on particularly advantageous terms:
 |
from a supplier which is definitely winding up its business activities |
 |
from the receivers or liquidators of a bankruptcy, an arrangement with creditors or a similar procedure. |
j) All tenders submitted in reply to an open procedure, a restricted procedure or competitive dialogue were irregular or
unacceptable. Only those tenderers were included in the negotiations which have satisfied the qualitative selection criteria.
Other justification for the award of the contract without prior publication of a
contract notice in the OJEU:
k) The contract has as its object services listed in Annex B of Directive 2004/18/EC or in Annex XVII B of Directive 2004/17/EC
l) The contract falls outside the scope of application of the relevant directive.
In addition to the box(es) ticked above please explain in a clear and
comprehensive manner why the award of the contract without prior publication of
a contract notice in the OJEU is lawful, stating in any case the relevant facts,
and, as appropriate, the conclusions of law in accordance with the articles in
the relevant Directive (500 words maximum):
The Francis Crick Institute intends to procure 2 Flow Cytometers from Becton Dickinson, as we are unaware of any other supplier who would be able to provide 2 high end 5 laser Flow-Cytometers, a sorter and an analyser, where both cytometers are identical in optical layout, optical filter configurations and fully compatible software, along with the patented Becton Dickinson technology. This would enable the Crick to analyse and sort samples containing Dangerous Pathogens (ACDP).
The instruments would installed to customised requirements as both instruments need to be contained within a custom-built Class I Biological Safety Cabinet which will require daily fumigation with no damage to laser or optical systems. Sufficient room is needed within the hood to manipulate samples safely so the systems must be capable of being modified to fit the required space constraints if necessary.
The key requirements for both cytometers are: Optical pathway:
— Reflective octagon and trigon-shaped detection system to maximize signal sensitivity on all parameters.
— Detection of a minimum of eighteen fluorescence parameters plus two light scatter parameters.
— Interchangeable optical filters without the need to realign.
Processing rate
— The cytometer will be capable of achieving processing/acquisition rates of 70 000 (seventy thousand) events per second at 18-bit resolution or higher.
Sensitivity and detection system
— Fluorescence Sensitivity measured at 70 psi and 488nm laser excitation will be less than 90 (ninety) MESF (molecules of equivalent soluble fluorochrome) for FITC and less than 30 MESF for PE channels and have a relative fluorescence detection efficiency (Qr) of more than 90 for FITC and more than 650 for PE.
— FSC resolution should be less than 0.6 microns.
— PMTs will have high signal to noise ratio, and be able to detect the entire fluorescence spectrum in the range of 300nm to 900nm.
Laser Lines
— 5 pinholes, one each for solid state laser operating at wavelengths around 355nm (three five five), 405nm (four zero five), 488nm (four eight eight), 561nm (five six one), 640nm (six four zero).
— Lasers should be aligned independently through separate pinholes — lasers should be spatially separated not co-linear.
— Lasers will be controlled independently via a software interface.
Software
— Both instruments will be operated using identical software so the same protocol can be used on either.
— Identical configuration and parameters will allow analysis experiments to be conducted on either instrument.
— QC will be monitored daily using the unique CS&T software on both instruments so they meet strict performance requirements.
Additionally, the sorter requirements are:
— Quartz cuvette flow cell coupled to the fluorescence objective lens by refractive index-matching optical gel for optimal collection efficiency and sensitivity.
— The cell sorter quartz cuvette flow cell will incorporate an integrated nozzle for fixed stream alignment.
— Sort Purity at 25 000 (twenty-five thousand) events per second will be over 98 %.
— Analysis or droplet deflection sorting will take place inside a fully enclosed chamber that will additionally prevent atmospheric air contamination by pathogenic materials.
— The synergistic hood and instrument design will enable daily fumigation of the flow cell to take place without damage to the laser and optical systems.
The Crick intends to enter into a contract with Becton Dickinson to procure the two Flow Cytometers described 10 days after the publication of this notice.
|